Water leak Detection | Gas leak detection | Pipe and Cable location | Object Detection

EX-TEC® GM 4 Download PDF

Gas warning and gas measuring instrument for detecting toxic gases,flammable gases, and oxygen

The ideal measuring instrument for gas and water suppliers, fire-brigades, sewage plants, landfills, biological gas plants and chemical industry.

EX-TEC® GM 4 Applications:
  • leak detection at pipes
  • monitoring environmental air
  • gas measuring to monitor chemical or biological processes


Features:
  • modular instrument concept, basic device without pump (diffusion device) or with built-in pump
  • electro-chemical sensors
  • interchangeable sensors by the operator
  • all commercial Mignon accummulators or batteries may be used for power supply
  • micro-processor controlled
  • large, illuminated LCD
  • self-test when switched-on
  • simple handling by three keys
  • internal data storage (PC evaluation possible)
  • visual and audible alarms
  • adjustable thresholds
  • temperature compensated sensors
  • user menu
  • function menu for individal instrument configuration (PIN-Code protected)
  • adjustable test gas concentrations


Sensors:

- Measuring range Response time t90 Durability
- carbon monoxide (CO) 0 ... 500 ppm < 30 s 2-3 years
- hydrogen (H2) 0 ... 10.000 ppm < 70 s 1-2 years
- hydrogen sulphide (H2S) 0 ... 100 ppm < 30 s 2-3 years
- hydrogen sulphide (H2S) 0 ... 2.000 ppm < 90 s 2 years
- ammonia (NH3) 0 ... 100 ppm < 60 s 1-2 years
- oxygen (O2) 0 ... 25 vol.% < 30 s 2 years
- hydrochloric acid (HCI) 0 ... 30 ppm < 70 s 1-2 years


Technical data

operating time: typ. 50 h diffusion operation; typ. 20 h pump operation
power supply: NiMH accu, rechargeable or 3 Mignon Alkaline batteries
EX-protection (CENELEC) EC-type-examination certificate according to ATEX 100 a,
guideline 94 / 9 / EG
- certificate No.: TÜV 01 ATEX 1657
- indification 1: Ex|| 2 G EEx e ib ||B T4, without leather bag
- indification 2: Ex|| 2 G EEx e ib ||C T4, with leather bag (for hydrogen)
pump capacity: -
- low pressure: > 150 mbar
- volumentric flow: typ. 5 ... 15 l/h
protection type: IP 54
operating temperature: -20 °C ... +40 °C
storage temperature: -20 °C ... +40 °C
humidity range: 15 ... 90 % r.h. , non-condensing
dimensions (WxHxD): 60x144x35 mm
weight: approx. 300 g


Accessories

  • GasIS (PC evaluation software) for documentation of measurements
  • charging technique for 12 V, 24 V or 100-240 V
  • leather bag
  • various probe systems
  • test set and test gases
  • case
EX-TEC® GM 4